ResMed VPAP Auto 25 VPAP ST VPAP S Sleep Apnea Machine User Manual


 
9Using the VPAP Auto 25, VPAP ST & VPAP S
English
Using the Data Card
If your clinician needs to review your treatment, they will ask you to use the Data Card
to copy data from your VPAP device and return the card to them.
Copying Data onto a Data Card
1 Switch on your VPAP and wait until you see the standby (Ramp/Settle) screen.
2 Hold the Data Card with the arrow facing up and insert it into the Data Card slot until
it stops. Data copying starts automatically.
The “Card Inserted Please Wait” message is displayed on the LCD while data is being
copied. Copying takes up to 30 seconds.
The “Copy Complete Remove Card” message is displayed on the LCD when copying
has finished.
3 Remove the Data Card by gripping the end of the Data Card and pulling it out.
4 Store the Data Card in its protective folder when not in use.
5 Return the card in its protective folder to your clinician using a postal envelope.
Updating Settings on your VPAP
If your clinician has provided a Data Card with new device settings:
1 With the device in standby (Ramp/Settle) mode, insert the Data Card into the slot on
the Data Card module. Updating will start automatically.
The “Card Inserted Please Wait” message is displayed on the LCD while updating is
in progress. Updating takes approximately five seconds.
The “Settings Success Remove Card” message is displayed on the LCD if the
settings were updated successfully.
Note: This message only appears once. If you re-insert the Data Card after you have
updated your settings, this message is not displayed.
2 Remove the Data Card from the VPAP device.
3 Store the Data Card in its protective folder when not in use.
!
WARNING
If your clinician has told you to use the Data Card to update the settings on your
device and the “Settings Success” message does not appear, contact your
clinician immediately.