ResMed S8 Series II Sleep Apnea Machine User Manual


 
13Technical Specifications
English
Technical Specifications
.
Note: The manufacturer reserves the right to change these specifications without notice.
Symbols which appear on the device
Attention, consult accompanying documents; Drip proof; Type CF equipment;
Dangerous voltage; Class II equipment; Start/Stop.
Operating pressure
range
4 to 20 cm H
2
O
Dimensions (H x W x D) 4.6” x 6.5” x 5.7” (112 mm x 164 mm x 145 mm)
Weight 2.9lb (1.3 kg)
Power Supply Input range for S8 Escape II Auto with H4i:100–240V, 50/60Hz; 110V,
400Hz; 2.5A < 140VA (110W) (maximum power consumption).
Instantaneous peak power consumption <340VA.
Refer to the DC-12 converter instructions for DC ratings. Only use
the ResMed DC-12 converter for DC input.
Typical power consumption at 20 cm H
2
0 is 34VA or 20W.
Air Filter Two-layered, powder-bonded, polyester non-woven fiber
Air Tubing Flexible plastic, 6’6” (2 m)
Air Outlet The 22 mm conical outlet complies with EN 1281-1
IEC 60601-1
Classifications
Class II (double insulation), Type CF
Sound Pressure Level Radiated sound pressure is measured at <26±2 dBA at 10 cm H
2
O and
3’3.36” (1 m) according to ISO17510-1.
Supplemental Oxygen Recommended maximum supplemental oxygen flow: 4 L/min
Housing Construction Flame retardant engineering thermoplastic
Operating Temperature +41ºF to +95ºF (+5ºC to +35ºC)
Operating Humidity 10%–95% non-condensing
Storage and Transport
Temperatur e
-4ºF to +140ºF (-20ºC to +60ºC)
Storage and Transport
Humidity
10%–95% non-condensing
Operating Altitude Sea level to 8,500’ (2,591 m)
Electromagnetic
Compatibility
Product complies with all applicable electromagnetic compatibility
requirements (EMC) according to IEC60601-1-2, for residential,
commercial, and light industry environments.
The electromagnetic compatibility tables for these ResMed devices can
be found on www.resmed.com, on the Products page under Service
& Support. Click on the PDF file for your device.