S8 ESCAPE
Performance: Operating pressure range: 4 to 20 cm H
2
O
Dimensions (H x W x D): 4.6 x 6.5 x 5.7 (112 mm x 164 mm x 145 mm)
Weight: 2.9lb (1.3 kg)
Power Supply
Input range for S8 ESCAPE with HUMIDAIRE 3i:100–240V, 50/60Hz; 110V, 400Hz;
2.5A < 140VA (110W) (maximum power consumption). Instantaneous peak
power consumption <340 VA.
Refer to the DC-12 converter instructions for DC ratings. Only use the ResMed
DC-12 converter for DC input.
Actual power consumption will vary, depending on factors such as the use of
accessories, height above sea level, and ambient temperature. Power consumption
values for typical treatment conditions (using a breathing machine set to 0.5 L and
15 breaths/min with 2 m (6’6”) air tubing and an
ULTRA MIRAGE mask) are given in the
following table.
Power outlet for HumidAire 3i
Outlet range: 100–240V, 50/60Hz; 110V, 400Hz; 2.0A
Air Filter: Two-layered, powder-bonded, polyester non-woven fiber
Air Tubing: Flexible plastic, 6’6” (2 m)
Air Outlet: The 22 mm conical outlet complies with EN 1281-1
IEC 60601-1 Classifications: Class II (double insulation), Type CF
Noise Level: Radiated sound pressure is measured at <30dBA at 10 cm H
2
O and
3’3.36” (1 m) according to ISO17510-1.
ACCESSORIES
DATA CARD
Dimensions (L x W x D): 2.56” x 0.87” x 0.03” (65.04 mm x 22.15 mm x 0.85 mm)
Device
Pressure
(cm H
2
O)
Power
(VA)
Power
(W)
S8 E
SCAPE
5178
10 23 12
15 29 16
20 34 20