ResMed Oxygen Equipment Oxygen Equipment User Manual


 
ENGLISH
TECHNICAL SPECIFICATIONS
PRESSURE / FLOW CURVE
Note: As a result of manufacturing variations, the vent flow may vary from the values listed in
the following table.
F
LOW GENERATOR SETTING
Select "Mir Full" as the mask option when using the ResMed Hospital Full Face Mask with
ResMed flow generators that have mask setting options.
DEAD SPACE INFORMATION
Explanation: Dead space is the empty volume of the mask up to the elbow.
The dead space of the mask varies according to cushion sizes but is less than 317 mL.
E
NVIRONMENTAL CONDITIONS
Operating Temperature: 5°C (41°F) to 40°C (104°F)
Operating Humidity: 15–95% relative humidity
SYMBOLS WHICH APPEAR ON THE DEVICE
Attention, consult accompanying documents.
ResMed Mask
S: Small; M: Medium; L: Large
REPLACEMENT PARTS
Pressure
(cm H
2
O)
Flow
(L/min)
422
832
12 41
16 48
20 54
Part
No.
Item
60700 ResMed Hospital Full Face Mask System, Small
60701 ResMed Hospital Full Face Mask System, Medium
60702 ResMed Hospital Full Face Mask System, Large
60710 Headgear (10pk.)
0
10
20
30
40
50
60
70
80
468101214161820
Mask Pressure (cm H
2
0)
0
10
20
30
40
50
60
70
80
468101214161820
Mask Pressure (cm H
2
0)
Flow Rate (L/min)
60867r3.book Page 5 Thursday, December 14, 2006 4:56 PM