Omron Healthcare NE-U22V Nebulizer User Manual


 
20
TROUBLESHOOTING GUIDE
PROBLEM CAUSE SOLUTION
The power indicator
does not illuminate.
The AC Adapter is not plugged
into an electrical outlet.
The connection stand is not
attached to the Main Unit.
The batteries have not been
inserted properly. The batteries
are low in charge.
Turn the power switch off. Plug the power
plug into an electrical outlet. Turn the device on.
Make sure Main Unit has been clicked onto
the Connection Stand.
Reinsert the batteries.
Replace both worn batteries immediately.
Recharge NiMH batteries with a
commercially available charger.
Residual medication has
dried on the components
and accessories.
Clean and disinfect the components
and accessories.
The mesh cap needs to be replaced.
Replace the Mesh Cap.
The power indicator
is illuminated but the
unit does not nebulize.
The batteries are low
in charge.
Replace same type batteries with new
alkaline or charged NiMH batteries.
The Medication Bottle has
too much medication in it.
Fill the Medication Bottle with the
proper amount of prescribed
medication. Max is 7 mL.
Liquid may have collected around
the electrodes of the Main Unit.
Absorb any moisture with a soft cloth.
There is liquid on top of the
Mesh Cap.
Remove visible liquid with a soft cloth
and a very light touch so as not to
damage the mesh.
The medication has not
come in contact with the
nebulizing parts.
Slightly slant the Main Unit towards you
with the ON/OFF button pointing down.
NE-U22V_EN_FR_SP_Q6_6.3L.qxd 2/18/09 3:13 PM Page 20