How to Disinfect
16
How to Disinfect the Unit
How to Disinfect
1 Make a 0.1% diluted solution using benzalkonium
chloride, such as Osvan S and soak the unit for 10
minutes.
- As for the dilution method, follow the instruction manual of benzalkonium
chloride.
- Do not leave the unit soaked in the solution for a long time.
2 Remove the unit from the solution and rinse it with water.
3 Immediately dry the unit by wiping off water with new
gauze.
This unit can be easily cleaned and disinfected because the Medication Bottle
and the Mesh Cap can be removed from the Main Unit.
The unit can be disinfected with benzalkonium chloride.
- Disinfection at the medical facility for medical treatment must be done after
confirming the target bacteria and the effects of disinfection method.
Do not disinfect the unit with the
disinfection solution containing
sodium hypochlorite.
The unit may not be able to be
used after getting covewred with
rust.
The following parts can be cleaned by boiling.
Mesh Cap, Inhalation Mask, Mouthpiece, Mask Adapter
(1) Place the parts in a container filled with sufficient water, and bring to boil.
(2) The parts can be disinfected by boiling for 10 to 30 minutes.
(3) When disinfection is finished, immediately dry the parts by
wiping thoroughly with new gauze.
- Never dry the parts in the microwave oven.
- Never wipe dry with tissues or cloth. Dust or lint left on the
Mesh may affect nebulization.
Parts that cannot be disinfected by boiling
Medication Bottle, Main Unit, Main Unit Cover,
Battery Cover, AC Adapter, Storage Bag