Omron Healthcare HEM-907XL Blood Pressure Monitor User Manual


 
12 13
FEATURES AND FUNCTIONS
FEATURES AND FUNCTIONS
Functions setting
(1) Inflation level setting
AUTO (Automatic setting): Can be used when the SINGLE, AVG, or MANU Mode is selected.
The unit estimates the systolic blood pressure during inflation and inflates to a proper cuff pressure
(approximately 30-40 mmHg above the patient’s systolic pressure).
Manual level setting: Inflation level can be set manually between 100 and 280 mmHg. Set the
level to 30 to 40 mmHg higher than the expected systolic pressure.
• To set the P-SET to "AUTO," turn the P-SET Knob counterclockwise as far as it goes until you can hear a click.
• In the "AUTO" setting, inflation level may not be set automatically when the systolic blood pressure is more than
220 mmHg. Use the unit in the manual setting.
• If the cuff has not been inflated to the necessary level, it may be re-inflated automatically.
ON/OFF
A
V
G
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
HIDE
P
-
SET
100
AUTO
140 180
220
260
280
MODE
AVG.
MANU.
CHECK
SINGLE
ON/OFF
A
V
G
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
HIDE
P
-
SET
100
AUTO
140 180
220
260
280
MODE
AVG.
MANU.
CHECK
SINGLE
(2) Non-display function
Use to prohibit the display of measurement results. However,
the cuff pressure during measurement is displayed.
This function can be used in the SINGLE and AVG Modes.
By pushing the Button, display or non-display of status
is switched alternatingly.
HIDE
(3) Manual deflation control
Accelerate deflation by pushing the DEFLATION (deflation
control) / Measurement Result Display Switch Button during
deflation in the measurement by the MANU Mode.
With each push of the Button, cuff is deflated rapidly
in increments of 5 to 10 mmHg.
BPM
mmHg
PULSE
SYS
mmHg
DIA
AC
AC
Display Non-display
P
-
SET
HIDE
ON/OFF
A
V
G
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
HIDE
AC
P
-
SET
100
AUTO
140 180
220
260
280
MODE
AVG.
MANU.
CHECK
SINGLE
XL
ON/OFF
A
V
G
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
HIDE
AC
P
-
SET
100
AUTO
140 180
220
260
280
MODE
AVG.
MANU.
CHECK
SINGLE
XL
ON/OFF
A
V
G
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
HIDE
AC
P
-
SET
100
AUTO
140 180
220
260
280
MODE
AVG.
MANU.
CHECK
SINGLE
XL
List of Functions for each Mode
Inflation level set-
ting
Non-display func-
tion
Manual deflation
control
Single (SINGLE)
Mode
Average
(AVG) Mode
Auscultation
(MANU) Mode
Check
(CHECK) Mode
Function
Measurement
Mode
(4) AVG Function setting
You can set the number of measurements, the waiting time until the 1st measurement, and the
measurement interval for the AVG Mode.
Items to set Set value
F1 Number of measurements 2 times or 3 times
F2
Waiting time until the start of 1st measurement
0 sec, 3 min, 5 min, or 10 min.
F3 Measurement interval 5 sec, 30 sec, 1 min, or 2 min.
Note: The bold letters represent the factory-set values.
Procedure to change the set values
1) When the power is OFF, push the ON/OFF (power) Button
for more than three seconds while holding the START Button;
F1 is displayed.
2) Push the START Button and select the function to set from
F1 to F3. Each time you push the START Button, the func-
tions change in the order of .
3) Push the DEFLATION (deflation control)/Measurement
Result Display Switch Button and change the set values.
4) When the setting is finished, push the ON/OFF (power)
Button to turn off the power. The setting is changed.
min
sec
Function #
HEM-907 10/23/03 3:13 PM Page 12