Omron Healthcare HEM-907 Blood Pressure Monitor User Manual


 
21
Clean the blood pressure monitor,the cuff and the
cables using a cloth dampened with water,diluted
disinfectant alcohol or diluted mild detergent.
To disinfect, wipe the unit with a soft cloth squeezed
well after moistened with the following disinfectant.
Benzalkonium chloride 0.01 - 0.2 w/v%
Chlorhexizin 0.05 - 0.5 w/v%
Amphoteric surface active agent 0.01 - 0.2 w/v%
When cleaning or disinfecting the unit, do not wipe the
AC adapter.
Then wipe the unit with a soft dry cloth.
1.
2.
How to Clean the Unit after Use
Caution
Pull the AC adapter from the electric outlet when
cleaning the unit.
You may suffer electric shock.
After cleaning the unit, dry it well, then insert the
AC adapter to the electric outlet.
You may suffer electric shock.
Requests from OMRON
Do not clean the unit with gasoline, thinner, or high
concentration alcohol.
Do not sterilize the unit by autoclave or gas sterilization
(EOG, formaldehyde, or high concentration ozone.)