Omron NE U07 Respiratory Product User Manual


 
14 15
English
Operation button
lamp (AC) does
not light up.
With AC
adaptor in use.
No nebulisation
Low nebulisation
rate
Symptom
Correctly insert it.
Correctly mount it.
Fill water up to proper level.
Fill the medication cup and switch
on the power by pressing OFF/ON
().
Replace the medication cup.
Remove the exess medication.
Increase delivery rate.
Go to AC operation.
Eventually read information sup-
plied with medication or consult
your doctor.
Remedy
If you find unsolvable problems please contact your authorised OMRON Customer Service.
For address see packaging and/or attached literature.
Check that the plug of the AC
adaptor is inserted correctly.
Check that the main unit is
mounted correctly onto the AC
adaptor.
Check that the level of water in
the water tank is normal.
Check that standard volume of
the medication is in the cup.
Check
Possible problems and how to solve them Technical data
This device fulfills the provisions of the EC directive 93/42/EEC (Medical Device Directive).
Model: OMRON NE U07
Electrical rating: AC 220V – 230V~, 50/60 Hz
Power consumption: Approx. 20 Watts
Particle size: 3.5 µm (MMD *)
Ultrasonic frequency: 2.4 MHz
Nebulisation rate: Approx. 1.0 ml/min.
Water tank: Approx. 21 ml
Fuse: = T 0.1 A
Medication cup: maximum 10 ml
Dimensions: Approx. 62 mm (W) x 110 mm (D) x 254 mm (H)
Weight: approx. 940 g
Operating conditions: 0 °C – 40 °C; 30% RH – 85% RH
Storage conditions: 0 °C – 40 °C; 30% RH – 85% RH
Electrical safety: - Class II,
- Type B
Symbol explanation: = Manual operation
AC = Alternating current
= OFF/ON
Technical data subject to change
Manufacturer:
OMRON HEALTHCARE Europe
* Mass Median Diameter
Data independently obtained by:
Malvern Instruments GmbH, Germany
on Malvern Master Sizer System
Be sure the medication cup is
not broken.
Check the medication level in
cup.
Check aerosol volume adjust
knob.
Check battery light.
Check preparation of
medication.