Omron M1 Plus Blood Pressure Monitor User Manual


 
35
8.Technical Data
EN
8. Technical Data
Note: Subject to technical modification without prior notice
This OMRON product is produced under the strict quality system of OMRON Healthcare
Co. Ltd., Japan. The Core component for OMRON blood pressure monitors, which is the
Pressure Sensor, is produced in Japan for assembly.
Disposal of this product and used batteries should be carried out in accordance
with the national regulations for the disposal of electronic products.
This device fulfils the previsions of EC directive 93/42/EEC (Medical Device
Directive). This blood pressure monitor is designed according to the European Standard
EN1060, Non-invasive sphygmomanometers Part 1: General Requirements and Part 3:
Supplementary requirements for electromechanical blood pressure measuring systems.
Caution: Please read the instruction manual carefully
before using the device.
Product Description Manual Inflation Blood Pressure Monitor
Model OMRON M1 Plus (HEM-4011C-E)
Display LCD Digital Display
Measurement Method Oscillometric method
Measurement Range
Pressure: 0 mmHg to 299 mmHg
Pulse: 40 to 180/min.
Memory 21 Measurements with date and time
Accuracy
Pressure: r3 mmHg
Pulse: r 5% of display reading
Inflation Manual by inflation bulb
Deflation Automatic pressure release valve
Pressure Detection Capacitive pressure sensor
Power Source 4 “AAA” batteries 1.5V
Battery life
Capacity of new alkaline batteries is approx. 1500
measurements
Operating temperature/
Humidity
+10°C to +40°C
Maximum: 30 to 85% RH
Storage temperature
Humidity
-20°C to +60°C
Maximum: 10 to 95% RH
Console Weight Approximately 210g without batteries
Cuff Weight Approximately 130g
Outer Dimensions Approximately 106 (w) mm u 80 (h) mm u 136 (l) mm
Cuff Dimensions
Approximately 146 mm u 446 mm
(Medium cuff: arm circumference 22 to 32 cm)
Package Content
Medium cuff, instruction manual, storage case, battery set,
guarantee card, blood pressure pass, regular bulb.
= Type B