Lumiscope 6700 Respiratory Product User Manual


 
16 6700-INS-LAB-RevC14
STORAGE
Keep the device and accessories dry. Avoid exposure
to direct sunlight, high moisture, and dust.
s NOTICE: If the device is to be stored long-term,
remove the battery.
s NOTICE: The battery's storage environment must be
between -4°F and 95°F (-20°C and 35°C).
s NOTICE: The device's storage environment must be
between -13°F and 140°F (-25°C and 60°C) and 10%
- 95% RH.
SAFETY PROTECTIONS
The nebulizer is equipped with protection against
several unsafe situations. If these situations occur, the
nebulizer will either (1) not power up or (2) turn off
automatically, as described below.
Protection against incorrect Power Supply
If an AC/DC adapter is used to power the device,
ensure the adapter meets local power regulations; if
not, the device may not power up (the Power Indicator
Light will not illuminate).
Protection Against Overheating
During normal operation, the Power Indicator Light will
illuminate GREEN. However, in the two following cases,
the device will indicate irregularity by turning off
automatically. The device then requires approximately
45 minutes at rest before restarting.
Protection Against Overuse: The electronic
components of the device are designed to function
continuously for approximately 40 minutes. Operating
the device for too long will trigger the heat-protection
mechanism and the device will turn off automatically.
Lack of Liquid in the Reservoir: In the event of lack or
insufficient liquid in the reservoir, the device will turn
off automatically.