Linear PERS-2400B Medical Alarms User Manual


 
4
13. TESTING TRANSMITTERS
15. LOCAL PROGRAMMING WITH RA-2400
14. ERASING TRANSMITTERS
16. TELEPHONE CONNECTIONS
PRESS AND HOLD TEST BUTTON FOR 5 SECONDS
• 6 BEEPS SOUND (ALL TRANSMITTERS WERE ERASED)
1
CAUTION: ALL TRANSMITTERS
WILL BE ERASED!
PRESS CLEAR TO RETURN TO OPERATE MODE
• 2 BEEPS SOUND WITH ACTIVITY TIMER ENABLED
• 3 BEEPS SOUND WITH ACTIVITY TIMER DISABLED
THE CONSOLE WILL RETURN TO OPERATE MODE
AUTOMATICALLY AFTER 3 MINUTES
2
PRESS TEST BUTTON
MOMENTARILY
• 4 BEEPS SOUND
(TEST MODE)
NOTE: CONSOLE WILL EXIT TEST MODE
AUTOMATICALLY AFTER 3 MINUTES
OR BY PRESSING THE CLEAR BUTTON
2
3
WHEN RA-2400 CONNECTION
TIMER STARTS COUNTING DOWN,
PRESS THE EMERGENCY BUTTON
• 4 BEEPS SOUND
(RA-2400 ANSWER)
1
TO LOCALLY PROGRAM
THE CONSOLE, CONNECT
THE MODEM'S LINE JACK
TO THE CONSOLE'S LINE
JACK
4
PERFORM THE STEPS DETAILED IN THE
RA-2400/PERS-2400 REFERENCE GUIDE (P/N 217163)
TO SET UP THE RA-2400 PROGRAM
LINE-TO-LINE
5
PERFORM PROGRAMMING IN
THE RA-2400 SOFTWARE, THEN
DISCONNECT FROM THE CONSOLE
CONNECT TELEPHONE
LINE TO THIS JACK
CONNECT LOCAL
TELEPHONE TO
THIS JACK
TELEPHONE
JACK
LOCAL
TELEPHONE
ROUTE CABLES UNDER
STRAIN-RELIEF TABS
CONSOLE WITH
REAR COVER
INSTALLED
3
2
1
PRESS TEST BUTTON
• 4 BEEPS SOUND (TEST MODE)
ACTIVATE EACH OF THE
CONSOLE'S TRANSMITTERS,
ONE AT A TIME
COUNT THE NUMBER OF BEEPS
THAT THE CONSOLE SOUNDS
WHEN THE TRANSMITTER
IS ACTIVATED
NOTE: WITH THE MODEL VB-4B VOICE BOARD
INSTALLED, THE CONSOLE WILL ANNOUNCE
THE TRANSMITTER STATUS
PRESS CLEAR TO RETURN TO OPERATE MODE
• 2 BEEPS SOUND WITH ACTIVITY TIMER ENABLED
• 3 BEEPS SOUND WITH ACTIVITY TIMER DISABLED
THE CONSOLE WILL RETURN TO OPERATE MODE
AUTOMATICALLY AFTER 3 MINUTES
4
1-BEEP = TX OK
2-BEEPS = TX HAD ALARM
3-BEEPS = TX MISSED STATUS
4-BEEPS = TX HAS LOW BATTERY