21
6. Assembly
6.2 Condenser
• Screw the condenser head into the condenser.
• Using the condenser height adjuster (5.4), turn
the condenser holder (5.1) downward as far
as it will go.
• Unscrew the clamping screw for the con-
denser (5.3) far enough so that the condenser
can be inserted from the front.
• From the front, insert the condenser into the
condenser holder as far as it will go. On the
underside of the condenser, there is an orien-
tation pin (6.1) that must be locked into place
in the guiding notch (7.1).
• Tighten the condenser’s clamping screw (5.3)
until the condenser locks into place.
• Connect the condenser over the connection
(8.1) with the stand.
Fig. 5 Condenser holder
1 Condenser holder
2 Condenser centering
3 Clamping screw for the condenser
4 Condenser height adjuster
Fig. 6
Underside of condenser
1 Orientation pin
Fig. 7 Condenser holder
1 Guiding groove
1
1
Fig. 8 Condenser connection
1 Condenser cable socket
Note:
The condenser must be centered before using
the microscope.
→
Köhler illumination p. 41.
1
23 4
1