Konica Minolta 300 Blood Pressure Monitor User Manual


 
9
Foreword
Difference from the SaO
2
value to displayed value
<Table 1> <Table 2>
Sa
O
2
HbCO
Sa
O
2
Methemoglobin
1% 5% 10% 1% 5% 10%
50 50
70 70
90 90
100 100
Measured value is alway
s
lower than the actual value.
Displayed value is sometimes
higher than the actual value if
Sa
O
2
is around 50%.
-0.1 -0.7 -1.5 +0.2 +1.3 +3.2
-0.2 -0.8 -1.6 -1.5 -6.0 -9.6
-0.1 -0.7 -1.5 -0.6 -2.3 -3.2
-0.2 -0.8 -1.7 -1.8 -7.5 -12.2
When influenced by abnormal hemoglobin like carbon
monoxide hemoglobin (HbCO) and methemoglobin
When pigments such as cardiogreen, intravascular dyes
and indocyanine greenare present in the blood
When fi nger nails are polished
When fi nger nails are polished
Tables 1 and 2 below show the errors which may occur due to
these hemoglobins.