Konica Minolta 300 Blood Pressure Monitor User Manual


 
8
Foreword
Notes on Use
This instrument should be used under the following operat-
ing conditions.
Temperature of 0 to 40°C (32 to 104°F), relative humidity
of 30 to 85%, with no condensation.
Atmospheric pressure of 700 to 1060 hPa (altitude of
400 to 3000m).
This is a precision instrument. To avoid the possibility of it
being damaged, the instrument should not be dropped nor
should heavy objects be placed on top of it.
Do not expose this
instrument to rain and water.
This instrument has no alarm function. Do not use this in-
strument if an alarm function is necessary.
This instrument is designed for use on adults. When using
the instrument on infants or babies, please note that the de-
signed measurement accuracy cannot be guaranteed.
Do not use the instrument on a point where there is exces-
sive vibration. In addition, to avoid excessive impact on the
instrument, it should be handled gently. Failure to observe
this may cause breakdown.
Accurate measurement data may not be obtained in the fol-
lowing cases.
When body movement is excessive
When the probe is not attached properly
When blood circulation at the measuring point is poor (due
to pressure on arm/finger, peripheral circulatory insuffi-
ciency)
When ambient light (e.g. panel light, fl uorescent light, in-
frared heat lamp, direct sunlight) is too strong
When infl uenced electromagnetically by other electronic
devices (e.g. near electrical appliances like TV, medial
device)
When a mobile telephone is used during measurement