Invacare XPO2 Oxygen Equipment User Manual


 
6
Invacare
®
Cylinders and Oxygen Delivery Device
Invacare
®
Cylinders and Oxygen Delivery Device
Cylinder Specifications
Model Number HF2POST6 HF2POST9 HF2POSTD IOH2M6 IOH2M9 IOH2MD
Height 12" 15" 21" 15.25" 14.25" 19.75"
Weight 4.2 lb. 5.4 lb. 6.5 lb. 3.5 lb. 4.8 lb. 7 lb.
Diameter 4.38" 4.38" 4.38" 3" 4.38" 4.38"
Capacity 170 Liters 255 Liters 422 Liters 148 Liters 248 Liters 422 Liters
Duration @ 2 lpm flow Varies Varies Varies 1 Hr 1.9 Hrs 3.5 Hrs
Fill Time 75 Min 125 Min 3.5 Hrs 1 Hr 1.9 Hrs 3.5 Hrs
Flow Settings N/A N/A N/A 0 - 6 lpm 0 - 6 lpm 0 - 6 lpm
Specifications
Battery usage time approx. 80 hours at 20 bpm
Flow rate range 0-6 lpm
Oxygen source pressure 20 psig (± 3 psi)
Regulator
Flow settings 0-6 lpm
Regulated pressure 20 psig
Length 5.4"
Thickness 1.6"
Width 3.6"
Weight 16 oz.
Model Selection Guide
Model no. Units/Case Limited warranty Description
IDD20EX 1 3 years 20 psi DODD with connecting tube
IRC2121 1 6 years 20 psi piston regulator, 0-6 lpm
2002919 1 1 year DODD pouch (attaches on cylinders)
M6/M9 BAG 1 1 year Shoulder bag, universal (M6/M9)
Invacare
®
Demand Oxygen Delivery Device
Model no. IDD20EX
Invacare’s Venture
®
Demand Oxygen Delivery Device (DODD)
is designed to give patients the freedom to do more of what
they want to do. When used with smaller cylinders, it is
lightweight, portable and extremely easy to carry. Extended
oxygen usage time makes it perfect for out-of-home use.
Through a wide range of respiratory rates (10-40 bpm), the
Venture DODD is able to deliver precise amounts of oxygen
at the very beginning of each breath in which delivery duration
is dependent on respiratory rate (variable-pulse technology).
And its alarms are clearly identified, making solutions simple
and quick.
Features
Variable demand will provide up to
4:1 conservation ratio
Internal clinicians switch will provide
up to 6:1 conservation ratio
Simple operation using only 2
buttons (on and off)
Uses one D-size disposable battery
Switches easily from continuous to
pulse modes; no disconnection
required
Includes small and lightweight pack
systems for easy carrying
Provides easy-to-use regulator for
setting flow
Cylinder Specifications
Model Number HF2PCM2KIT HF2PCL4 HF2PCL6 HF2PC9
Height 8.25" 9" 11" 14"
Weight 2 lb. 3.1 lb. 3.5 lb. 4.8 lb
Diameter 3" 4.38" 4.38" 4.38"
Capacity 57 Liters 113 Liters 170 Liters 255 Liters
Conserved Durations (Hrs) @ 20 bpm
At setting of 1 3.4 Hrs 6.7 Hrs 9.8 Hrs 15.2 Hrs
At setting of 2 2.2 Hrs 4.3 Hrs 6.3 Hrs 10.5 Hrs
At setting of 3 1.6 Hrs 3.2 Hrs 4.7 Hrs 7.3 Hrs
At setting of 4 1.3 Hrs 2.6 Hrs 3.8 Hrs 5.9 Hrs
At setting of 5 1.1 Hrs 2.1 Hrs 3.1 Hrs 4.8 Hrs
Fill Time 24 Min 55 Min 75 Min 125 Min
Flow Settings 1-5 Pulse 1-5 Pulse 1-5 Pulse 1-5 Pulse
or 2 lpm Cont. or 2 lpm Cont. or 2 lpm Cont. or 2 lpm Cont.
280274_05-054:05-054 1/21/08 12:36 PM Page 6