Invacare RLS6 Scale User Manual


 
Reliant Scale 4 Part No. 1114832 Rev B
INSTALLING THE RELIANT SCALE (FIGURE 2)
1. Position the load cell assembly of the Reliant Scale into the boom mounting bracket.
Refer to DETAIL "A" in FIGURE 2.
NOTE: For 9805 lift use 1/8-inch thick nylon washer. For 450 lb/600 lb lifts use 1/4-inch nylon
washer.
2. Secure the Reliant Scale to the boom mounting bracket with a shoulder bolt, one (1)
small steel washer, two (2) nylon washers and a locknut. Securely tighten. Refer to
DETAIL "A" in FIGURE 2 for washer orientation.
NOTE: Ensure the shaft of the shoulder bolt passes through both sides of the boom mounting
bracket.
3. Insert swivel bar pin with two (2) nylon washers through the swivel bar. Refer to DETAIL
"B" in FIGURE 2.
4. Align the mounting holes in the swivel bar pin with the mounting holes in the
load cell assembly. Refer to DETAIL "B" in FIGURE 2.
5. Secure swivel bar pin to the load cell assembly with the provided mounting screw and locknut.
Securely tighten. Refer to DETAIL "B" in FIGURE 2.
6. Weigh the patient. Refer to OPERATING THE SCALE in this instruction sheet.
FIGURE 2 - INSTALLING THE RELIANT SCALE
Boom
Boom
Mounting
Bracket
Locknut
Small
Steel
Washer
Nylon
Washer
Nylon
Washer
Shoulder
Bolt
Reliant
Scale
Load
Cell
Assembly
DETAIL "A"
Swivel Bar Pin
Reliant
Scale
Swivel Bar
Locknut
Mounting
Screw
DETAIL "B"
Nylon Washers