Invacare PLATINUM 5 Respiratory Product User Manual


 
17
NOTE: The use of some accessories such as the Homefill II compressor will
deactivate the Low Flow Alarm.
SENSO
2
OXYGEN PURITY INDICATOR
(IF YOUR UNIT FEATURES THE O
2
SENSOR FEATURE)
This feature monitors the purity of oxygen generated by the oxygen
concentrator. If purity falls below factory preset standards, indicator
lights on the control panel will illuminate.
INITIAL STARTUP OF THE CONCENTRATOR
NOTE: Concentrator may be used during the initial start warm-up time
(approximately 30 min.) while waiting for the O
2
purity to reach maximum.
When the unit is turned on, the green light will come on (SYSTEM OK / O
2
greater than
85%). After 5 minutes, the oxygen sensor will be operating
normally and will control the indicator lights depending on oxygen concentra-
tion values. The explanation of the indicator light functions are as follows.
WARNING
DO NOT change the L/min. setting on the flowmeter unless
a change has been prescribed by your physician or therapist.
NOTE: To properly read the flowmeter, locate the prescribed flowrate line on
the flowmeter. Next, turn the flow knob until the ball rises to the line. Now,
center the ball on the L/min. line prescribed (FIGURE 7).
2. If the flowrate on the flowmeter ever falls below 0.5 L/min. for
more than about one minute, the LOW FLOW ALARM will be
triggered. This is a rapid beeping of the audible alarm. Check your
tubing or accessories for blocked or kinked tubing or a defective
humidifier bottle. After rated flow is restored, more than
0.5 L/min., the Low Flow audible alarm will go off.
OPERATING INSTRUCTIONS
FIGURE 7 - ADJUSTING THE FLOWRATE-
PLATINUM 5 WITH SENSO
2
SHOWN
Ball
Flow Knob