Invacare PLATINUM 5 Respiratory Product User Manual


 
10
SPECIFICATIONS
SPECIFICATIONS (CONTINUED)
Label Status Indicator Lights (LED)
Symbol
I/O SYSTEM OKAY GREEN
SYSTEM FAILURE RED
5LX INDICATORS:
Flow Range:
Low Flow Alarm:
Average Power
Consumption:
Pressure Relief
Mechanism Operational at:
Change in maximum
recommended flow
when back pressure
of 7kPa is applied:
Filters:
Safety System:
Width:
Height:
Depth:
Weight:
Shipping Weight:
1/2 - 5 L/min.
Flowrates less than 1 L/min., are not
recommended.
0 L/min. to 0.5 L/min., Rapid Audible
Alarm Beeping (No Accessories Connected)
395W
241 kPa ± 24.1 kPa (35 psi ± 3.5 psi)
0.7 L/min.
Cabinet (2), Outlet HEPA and
Compressor Inlet.
Current overload or line surge shutdown.
High temperature compressor shutdown.
High Pressure Alarm w/compressor shutdown.
Low Pressure Alarm w/compressor shutdown.
Battery Free
Power Loss Alarm.
SensO
2
Oxygen System (SensO
2
Model)
Low Flow Alarm.
46.7 cm ± 1 cm (18 3/8 ± 3/8)
67.0 cm ± 1 cm (26 3/8 ± 3/8)
36.5 cm ± 1 cm (14 3/8 ± 3/8)
23.2 kg ± 0.5 kg (51 lbs. ± 1 lb)
26.3 kg ± 0.5 kg (58 lbs. ± 1 lb)