SECTION 6—MAINTENANCE
Platinum
™
Series 32 Part No 1118353
SECTION 6—MAINTENANCE
NOTE: The Invacare concentrators are specifically designed to
minimize routine preventive maintenance. In places with high dust
or soot levels, maintenance may need to be performed more often.
Only authorized or factory-trained personnel should perform
preventive maintenance on the concentrator. Power should be
disconnected before beginning preventive maintenance on the
Invacare concentrator.
NOTE: At a minimum, preventive maintenance MUST be
performed according to the maintenance record guidelines. Refer to
Preventive Maintenance Record on page 34.
Routine Maintenance
NOTE: The following routine maintenance should be performed
more frequently. See recommended intervals in each section and
preventive maintenance record.
Cleaning the Cabinet Filter
NOTE: For this procedure, refer to FIGURE 6.1.
NOTE: There are two cabinet filters one located on each side of the
cabinet.
1. Remove each filter and clean at least once a week
depending on environmental conditions.
NOTE: Environmental conditions that may require more frequent
cleaning of the filters include but are not limited to: high dust,
smoking, air pollutants, etc.
2. Clean the cabinet filters with a vacuum cleaner or wash in
warm soapy water and rinse thoroughly.
3. Dry the filters thoroughly before reinstallation.