SECTION 5—OPERATING INSTRUCTIONS
Platinum
Series 30 Part No 1118353
YELLOW light ( ) - Call supplier IMMEDIATELY. You may
continue to use the concentrator unless instructed otherwise
by your supplier. Be certain that backup oxygen is nearby.
RED light ( )- Total unit shutdown. Switch IMMEDIATELY
to backup oxygen supply. Call supplier IMMEDIATELY.
GREEN light - with YELLOW light flashing - Call supplier
IMMEDIATELY. Oxygen sensor malfunctioning; you may
continue to use the concentrator.
FIGURE 5.7 SensO
2
Oxygen Purity Indicator (If Your Unit
Features the O
2
Sensor Feature)
Indicators (If Your Unit Does Not
Feature the O
2
Sensor)
NOTE: For this procedure, refer to FIGURE 5.8 on page 31.
LX models indicator Light Explanation
RED light ( ) - Total Unit Shut-Down. Switch
IMMEDIATELY to a back-up oxygen supply. Call supplier
IMMEDIATELY.
GREEN light (I/O) - On/Off. System okay.
Call Supplier
Normal
Below Normal
OXYGEN
PURITY
RED
YELLOW
GREEN
O
2
Auto Shut - Down
Use Backup