Invacare IRC10LXO2 Respiratory Product User Manual


 
SECTION 7—TROUBLESHOOTING GUIDE
Platinum
Series 36 Part No 1118353
YELLOW or RED
LIGHT Illumi-
nated.
Alarm:
CONTINUOUS
On red light only.
* Only applies to
5LXO
2
and
10LXO
2
SensO
2
models.
1. Low oxygen purity.*
2. Kinked or blocked tub-
ing, cannula or humidi-
fier.*
3. Flowmeter set at 0.5
L/min. for Platinum 5 and
1 L/min. for Platinum 10
or less.*
4. Unit overheating due
to blocked air intake.
5. Insufficient power at
outlet.
6. Internal repairs
required.
1. Clean or Replace filters.
2. Inspect for kinks or block-
ages. Correct, clean or
replace item. Once cor-
rected, turn power OFF for
60 seconds and then turn
power back ON.
3. Check flowmeter is set to
1.0 L/min. or more. (Refer to
Recommended Guidelines for
Optimum Performance on
page 24.
4.a. Remove and clean cabinet
filters.
4b. Move concentrator at
least three inches from walls,
draperies, and furniture.
5. DO NOT use extension
cords. Move to another elec-
trical outlet or circuit.
6. Call Service Provider.
GREEN LIGHT
with YELLOW
Light Flashing.
1. Internal repairs
required.
1. Call Service Provider.
SYMPTOM: PROBABLE CAUSE: SOLUTION: