SECTION 7—TROUBLESHOOTING GUIDE
Part No 1118353 35 Platinum
Series
SECTION 7—TROUBLESHOOTING
GUIDE
SYMPTOM: PROBABLE CAUSE: SOLUTION:
Alarm: Main Power Loss: 1. Insert plug into outlet.
SHORT BEEPS,
LONG PAUSE
1. Power cord NOT
plugged in.
2. Inspect house circuit
breakers or fuses. If problem
recurs, use a different outlet.
Concentrator
NOT operating,
power switch
ON.
2. No power at outlet. 3. Call Service Provider.
Beep......Beep...... 3. Tripped circuit
breaker.
Alarm:
CONTINUOUS
Concentrator
NOT operating,
Power Switch
ON.
Beeeeeeeep....
System Failure:
1. Unit overheating due
to blocked air intake.
2. Insufficient power at
outlet.
3. Internal repairs
required
1a. Remove and clean cabinet
filters.
1b. Move oxygen concentra-
tor at least three inches away
from walls, draperies or furni-
ture.
2. DO NOT use extension
cords. Move to another elec-
trical outlet or circuit.
3. Call Service Provider.