SECTION 6—CYLINDER FILL TIMES
HomeFill
20 Part No 1148114
SECTION 6—CYLINDER FILL TIMES
NOTE: All filling times are approximate and may vary on
environmental conditions.
*
Troubleshooting
If post valve cylinder fails to function, try the following
solutions. If problem cannot be corrected, consult your
dealer/home care equipment provider.
Replacement Parts - Accessories
HF2POST6BAG or HF2POST9BAG Carrying Bag.
Cylinder
Filling
Times
Concentrator Flow Rate to Patient:
(Platinum 5 up to 3 L/min. and Platinum 10 up to 5
L/min)
HF2POST6 HF2POST9 HF2POSTD
NORMAL 1 hr 25 min 2 hr 20 min 3 hr 20 min
MAXIMUM 1 hr 50 min 2 hr 45 min 3 hr 50 min
PROBLEM POSSIBLE CAUSE SOLUTION
No flow. 1. Regulator or
conserver is in “Off”
position.
1. Set to prescribed
setting.
2. Cylinder empty. 2. Replace/fill cylinder.
3. Toggle Valve closed. 3. Open toggle valve.
4. Conserver not
sensing breath.
4a. Check position of
cannula in nose. Do not
breathe through mouth.
4b. Check to see if the
cannula is kinked.