Invacare HF2POST6 Oxygen Equipment User Manual


 
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
Part No 1148114 11 HomeFill
External Examination
NOTE: For this procedure, refer to FIGURE 4.1 on page 12.
1. Examine the outside of the cylinder for the following
conditions and replace the cylinder if they exist:
Dents or dings
Arc burns
Oil or grease
Any other signs of damage that might cause a cylinder
to be unacceptable or unsafe for use.
2. Examine the cylinder for evidence of fire or thermal
damage. Evidence includes charring or blistering of the
paint, or other protective coating or heat sensitive
indicator. If fire or thermal damage is found, replace the
cylinder.
3. Inspect the cylinder fillport for the following and replace
if found:
Debris, oil or grease
Noticeable signs of damage
Signs of corrosion inside the valve
Signs of excessive heat or fire damage