Invacare 9000 Series Mobility Aid User Manual


 
9000 Series 30 Part No. 1056953
Arm Socket (INSIDE of
Wheelchair Frame)
Hex Screw
Locknut
Left Side of Wheelchair Frame
FIGURE 5 - CONVERTING FROM SPACE SAVER TO CONVENTIONAL
ARMS OR CONVENTIONAL TO SPACE SAVER ARMS
(9000 XT/XDT WHEELCHAIRS)
Right Side of
Wheelchair Frame
SPACE SAVER ARM POSITION
CONVENTIONAL ARM POSITION
Hex Screw
Locknut
Arm Socket (OUTSIDE
of Wheelchair Frame)
Wheelchair
Frame
CONVERTING FROM SPACE SAVER TO CONVENTIONAL
ARMS OR CONVENTIONAL TO SPACE SAVER ARMS
(9000 XT/XDT WHEELCHAIRS) (FIGURE 5)
NOTE: Space Saver arms DO NOT work with the following:
A. 9000XT with the adjustable angle back option.
B. The 9000XT Recliner.
1. Remove the arms from the wheelchair. Refer to REMOVING OR REPLACING
ARMRESTS in this section of the manual.
2. Remove the locknut and hex screw that secure the arm socket to the wheelchair frame.
3. Repeat STEP 2 for the opposite side of the wheelchair.
4. Reposition the arm socket to one (1) of two (2) positions:
ARM TYPE ARM SOCKET POSITION (On wheelchair Frame)
Conventional OUTSIDE
Space Saver INSIDE
NOTE: Make sure arm sockets sit flush against the wheelchair frame.
5. Reinstall the hex screw and locknut that secure the arm socket to the wheelchair
frame and securely tighten.
6. Repeat STEP 5 for the opposite side of the wheelchair.
7. Install the NEW arms onto the wheelchair. Refer to REMOVING OR REPLACING
ARMRESTS in this section of the manual.
8. Refer to the following chart to determine if the rear wheel axle spacer is needed:
ARM TYPE AXLE SPACER NEEDED?
Conventional Arms YES
Space Saver Arms NO
9. If rear wheel axle spacer is needed, refer to REMOVING/INSTALLING THE REAR
WHEELS in SECTION 6 of this manual.
SECTION 4 ARMS
ARMS