Invacare 5 Oxygen Equipment User Manual


 
SECTION 5 —OPERATING INSTRUCTIONS
Part No 1145728 19 Platinum™Series
Recommended Guidelines for Optimum
Performance
NOTE: Ensure that your concentrator is at least three inches away
from walls, draperies or furniture to assure sufficient air flow. Avoid
deep pile carpets and heaters, radiators or hot air registers.
Plug in Power Cord
1. Plug in power cord to an electrical outlet.
Temperature: 50°F - 95°F (10°C - 35°C)
Electrical: No extension cords.
Placement: No closer than three inches from the wall,
furniture, draperies, or similar surfaces.
Tubing and Cannula: 7 ft cannula with a maximum 50 ft of
crush-proof Tubing (DO NOT pinch).
Platinum 10: Recommended use up to 50 ft
high flow tubing with high flow cannula at all
flow rates.
Environment: Smoke and soot-free. No confined spaces
(Example: No closets).
Relative Humidity: 20 to 60%
Time of Operation: Up to 24 hours per day.
Flow Rate Platinum 5 and XL: From 0.5 L/min. to 5
L/min. For flow rates less than 1 L/min., use
the Pediatric Flowmeter Accessory.
Platinum 10: From 2 L/min. to 10 L/min
Minimum Operating Time: 30 Minutes