SECTION 4 —SPECIFICATIONS
Part No 1145728 15 Platinum™Series
5LXO2 INDICATORS:
5LX INDICATORS:
Flow Range: Platinum 5 and XL - 0.5 to 5 L/min. For
flowrates less than 1 L/min., we recommend
the use of the Invacare Pediatric Flowmeter
IRCPF16 Accessory.
Platinum 10 - 2 L/min. to 10 L/min
Low Flow Alarm: Platinum 5 - 0 L/min. to 0.5 L/min.,
Platinum 10 - 0 L/min. to 1 L/min.
Rapid Audible Alarm Beeping (No
Accessories Connected)
Average Power
Consumption:
Platinum 5 and XL: 400W
Platinum 10: 585W
Pressure Relief Mechanism
Operational at:
35 psi ± 3.5 psi (241 kPa ± 24.1 kPa)
Change in maximum
recommended flow when back
pressure of 7kPa is applied:
0.7 L/min.
O
2
Purity
SYSTEM FAILURE
Below 73%
73% to 85%
SYSTEM OKAY
Over 85%
Indicator Lights
RED - Continuous Audible
Alarm Sieve-GARD
Compressor Shutdown
YELLOW
GREEN
O
2
Label
Symbol
Status
SYSTEM FAILURE
SYSTEM OKAY
Indicator Lights
RED
Label
Symbol
I/O
GREEN