FIGURE 2 - REMOVING THE FRAME FROM
THE BATHTUB
TILT FRAME
FORWARDS
Frame
Bathtub
Suction Cup
Loop
Suction
Cup
Self-Locking
Screws
(STEP 2)
Mounting Screw
(STEP 2)
Washer
(STEP 2)
Washers
(STEP 2)
Motor Shroud
(STEP 3)
FIGURE 3 - REMOVING THE MOTOR
SHROUD
CAUTION
The motor shroud is still connected to the
motor by the spade connectors and pen-
dant cable. DO NOT attempt to forcibly
separate the motor shroud from the mo-
tor. Otherwise - damage to the motor
shroud and/or motor may occur.
3. Carefully lift the motor shroud off of the motor.
4. Disconnect the RED and BLACK spade connec-
tor wires from the motor as shown in FIGURE 4.
5. Remove the two (2) mounting screws securing
the mounting bracket to the motor as shown in
FIGURE 4.
NOTE: Save the EXISTING two (2) mounting screws
for reuse when securing the NEW spade connector
wire and pendant cable to the motor.
6. Remove the mounting bracket and release the
spade connector wire and pendant cable as shown
in FIGURE 4.
7. Disconnect the battery connector from the motor
shroud. Refer to
REMOVING THE BATTERY
CONNECTOR in this instruction sheet.
Removing the Motor Shroud (FIGURES 3
AND 4)
WARNING
ALWAYS lower the bath lift frame to the
lowest position BEFORE performing this
procedure. Otherwise, the bath lift frame
may tip during installation resulting in in-
jury and/or damage.
WARNING
ALWAYS remove the bath lift from the bath-
tub BEFORE performing this procedure.
Otherwise - injury or damage may occur.
1. Remove the bath lift frame from the bathtub, if not
already done. Refer to
REMOVING THE FRAME
FROM THE BATHTUB in this instruction sheet.
2. Remove the mounting screw, two (2) self-locking
screws and three (3) washers securing the motor
shroud to the bath lift frame. Refer to FIGURE 3.
Removing the Frame from the Bathtub
(FIGURE 2)
WARNING
ALWAYS lower the bath lift frame to the
lowest position BEFORE performing this
procedure. Otherwise, the bath lift frame
may tip during installation resulting in in-
jury and/or damage.
CAUTION
ALL water MUST be drained from the bath-
tub BEFORE removal of the bath lift frame
to prevent damage to the bath lift frame
or suction cups.
ALWAYS release the suction cups from bath-
tub bottom BEFORE removing frame to pre-
vent damage to the frame or bathtub.
1. Remove the seat, if not already done. Refer to
RE-
MOVING THE SEAT in this instruction sheet.
NOTE: Ensure bath lift is in lowest position before pro-
ceeding.
2. Grasp the handgrip on the battery pack and pull
upwards to remove it from the motor shroud.
3. Tilt the frame forward towards the drain or gently
pull up on the suction cup loops to release the suc-
tion cups from the bathtub.
4. Remove the frame from the bathtub.
5. Remove the motor shroud. Refer to
REMOVING
THE MOTOR SHROUD in this instruction sheet.