IDEAL INDUSTRIES 61-635 Thermometer User Manual


 
Page 8
Required Equipment
Required equipment is listed in Table B. If the recommended
models are not available, equipment with equivalent specifications may be used.
Repairs or servicing should be performed only by qualified personnel.
Table B. Required Equipment
Equipment Required Characteristics
Recommended
Model
Calibrator AC Voltage Range: 0-750V ac
Accuracy: ±0.07% (Basic)
Frequency Range: 40 ~ 100KHz
Accuracy: ± 2%
DC Voltage Range: 0-1000V dc
Accuracy: ±0.006% (Basic)
Current Range: 0 ~ 10A
Accuracy:
AC (40Hz to 1KHz): ±0.08% (Basic)
DC: ± 0.02% (Basic)
Frequency Source:
5.00Hz ~ 1.0000MHz
Accuracy: ±0.001%
Amplitude:
0.5V p-p ~ 1.0V p-p (square wave)
Accuracy: ± 5%
range: 1 ~ 2G
Accuracy: ±0.03% (Basic)
Capacitance Range: 1PF ~ 40mF
Accuracy: ±0.10% (Basic)
Temperature Range: -200 ~ 1200
Accuracy: ±0.3 (Basic)
Fluke 5500 or
Wavetek 9100
Calibrator or
equivalent
Form Number TM61633-5 Rev 5 Dec 2006