Health O Meter 2450KL Wheelchair User Manual


 
19
CALIBRATION PATH
The calibration is performed in Kg or Lb, according to the units of measure used upon entering
into the calibration path.
Operator Action Display
1. Press ON/OFF button and immediately press and
hold HOLD/RELEASE button for 3-4 seconds
Calibration
Process
2. Press ENTER Enter load weight
600.0
3. Using cd keys and/or the keypad, set the
calibration load to at least 200 lbs (for best results
and greater accuracy, use a 600 lb weight). Press
ENTER
Zero calibration
Please clear the scale
4. Please clear the scale platform and press ENTER Zero calibration
Please wait
5. Please wait without touching or shaking the scale
until the zero calibration process is finished
Weight calibration
Put: XXX.X
6. Load the scale platform with the required weight
and press ENTER
Weight calibration
Please wait
7. Please wait without touching or shaking the scale
until the calibration process is finished
Calibration
factor: X.XXXXX
8. Press ENTER Calibration
Please clear the scale
9. Please remove weight from the scale platform Rebooting
Please wait
10. Please wait until the scale resumes normal
operation