Health O Meter 2450KL Wheelchair User Manual


 
18
TROUBLESHOOTING
Refer to the following instructions to check and correct any failure before contacting service
personnel.
SYMPTOM POSSIBLE CAUSE CORRECTIVE ACTION
Scale does not turn on 1. Dead Battery
2. Faulty electrical outlet
3. Bad power supply
1. Replace batteries
2. Use a different outlet
3. Replace adapter
Questionable weight or
the scale does not zero
1. External object
interfering with the scale
1. Remove interfering object
from the scale
2. The display did not show
”0.0” before weighing
2. Ask the patient to get off
the scale, zero the scale
and begin weighing
process again
3. Scale is not placed on a
level floor
3. Place the scale on a level
floor and begin weighing
process again
4. Scale is out of calibration 4. Check weight with known
weight value
5. Improper Tare 5. Place the item to be tared
on the scale ramps. Press
REWEIGH. Once the
weight of the item is
displayed, press TARE.
Place the patient and the
tared item back on the
scale. Press REWEIGH
again.
Weighing is performed but
the display shows “weigh”
and “reweigh” every few
seconds; the weighing
process takes too long and
no weight is displayed.
The patient is not sitting still Ask the patient to be still or
you can change to live weight
setting
The display shows
“Overload” message
The load on the scale
exceeds the capacity
(600 Lbs)
Remove the excess weight
and use the scale according to
its limits
The display shows
“LOW BAT” message
The batteries are empty Replace batteries according to
instructions
The display shows
“Load Cell Error” message
There is a problem with one
or more load cells or the load
cell cable is disconnected.
Check load cell cable
connection at the display and
platform assembly ports. If the
problem is not corrected, refer
to qualified service personnel
to replace the defective load
cell