Graham Field LF1050 Personal Lift User Manual


 
LF1050-INS-LAB-RevE13
18
CLEANING
Regular cleaning is recommended every three months; clean the lift more often as needed. A soft
cloth, dampened with water and a mild NON-ABRASIVE detergent (household cleaner, soapy water,
or hospital grade disinfectant) is all that is needed to clean your patient lift. Automobile wax or
furniture polish will help maintain the nish over a long period of time.
s NOTICE: DO NOT wash patient lift under water pressure or steam clean.
CARE OF SLINGS
Lumex slings are washable at 150° F and autoclavable at 185° F for 30 minutes. The head support’s
plastic insert should be removed before washing. The slings can be air dried or dried by using a
dryer on LOW heat.
s NOTICE: DO NOT bleach slings!
MAINTENANCE SCHEDULE
Info: Only qualified persons should service and repair your Lumex LF1050 Patient Lift.
Regular maintenance of your patient lift is necessary to ensure continuing proper and safe
operation. Read and observe the following recommended maintenance schedule:
Item Inspect Inspect / adjust
initially monthly every 3
months
Boom & spreader bar
Checkconnectionsbetween1)boomandspreaderbarand2)boomandmast
forimproperconnection,looseness,orwear
X X
Inspectboomforbendingordeection X X
Mast
Inspectmastforbendingordeection X X
Inspectsteeringbarfordamageorloosehardware X X
Base and foot pedal
Ensureallhardwareistight X X
Ensurecastersandaxleboltsaretight X X
Inspectrubberpartsofcastersfordeection X X
Applygreasetocasterballbearingsifneeded X
Actuator and control box
Ensurecontrolboxisrmlyafxedtothemast X X
Ensurepinsthatconnectactuatortoboomandmastarermlyafxed X X
Ensurecablethatconnectsactuatorandcontrolboxisrmlyafxed X X
Ensurebatteryindicatorhasatleast2greenLEDsilluminatedbeforeuse X X
Sling and sling hardware
Checkslingforwear;discardifworn X X
Ensureslinghardwareisingoodconditionbeforeeachuse X X