8 TROUBLESHOOTING
WARNING: If you experience a problem with your lift and are unable to service it your-
self, contact your Lumex dealer.
SYMPTOM POSSIBLECAUSE ACTION
Patientliftdoesnotgoup
ordown
Airinthecylinder Bleedthecylinderbyturningtherelease
knobandpushdownonthemast.Pump
uptheliftagainandturnthereleaseknob
again.
Lossofhydraulicuid(leaking
cylinder)
Replaceanyleakingcylinder
Patientliftdoesnotgo
down
Airinthecylinder SeeaboveDirections
Info: If the hydraulic cylinder begins leaking fluid, do not operate. Contact your Lumex
dealer for a new cylinder.