Graham Field LF1031FP Personal Lift User Manual


 
1031FP-INS-LAB-RevA11
17
Installation of New Pump
Info: The hardware for attaching the Hydrau-
lic Pump (GF #1010S005) to the mast and
boom comes with the lift frame.
1. Attach the bottom of the new hydraulic pump
to the lower mast bracket with the pressure
knob and pump lever toward the bottom of
the pump, as shown at near right, and the
FRONT” label toward you, as shown at far
right. Use a hex bolt, washers, and hex nut to
secure pump to mast. Use two
9
/16" wrenches
to tighten.
2. Attach the top of the hydraulic pump to the
upper boom bracket, as shown at right. Use
a hex bolt, washers, and hex nut to secure
pump to boom. Use two
9
/16" wrenches to
tighten.
WARNING: Ensure that the Hydraulic Pump
is installed as described. If the Hydraulic
Pump is not properly installed, personal
injury and damage to the Hydraulic Pump
could result.
boom
upper boom
bracket
attach new pump
to upper boom
mast
attach new pump
to lower mast
detail — attach new
pump to lower mast