c. To change the default SET TEMP : You can adjust this
temperature within the range of ambient to + 56 °C, in one
degree increments. To adjust, press the SET TEMP key - then
press the numeric key pad to indicate the temperature desired.
All keys must be pressed within 5 seconds of each other. The
display should now show the new SET TEMP.
When the SET TEMP is same as the WATER TEMP, the
bath is ready to be used for thawing.
d. Place the frozen product in slot marked 1 on the Plasma Bag
Holder.
It is recommended that frozen plasma be placed into a plasma
overwrap thawing bag. This should prevent plasma from
getting into the bath, should the bag be cracked or damaged
and protect the device from damage. After placing frozen
plasma into the plasma thawing bag, eliminate all air from
bag, twist top of the bag and loop it through the Plasma Bag
Holder so it is secure.
For additional overwrap thawing bags, call
GenesisBPS, LLC.
It is important to remove the excess air. Place the
weight(s) provided into the plasma bag(s) to ensure that the
frozen plasma remains submerged in the water.
e. Press the START/CHANGE key for the selected bag location.
The timer will start it’s count down while the unit is thawing
the plasma. Default timing is 14 minutes for the GENESIS
2002 and 18 minutes for the GENESIS 2006. After this
time has elapsed an alarm will sound, signaling that the
frozen product should now be thawed.
Depending on volume of plasma, the thickness of the plasma
thawing bag and storage temperature of product to be thawed,
7