Fluke Food Safety Thermometer Thermometer User Manual


 
FoodPro Plus
Users Manual
2
FoodPro Plus Specifications and Features
See Table 1 for specifications and features. Specifications are subject to
change without notice.
Table 1. Specifications and Features
Infrared
Temperature Range IR Channel -35 to +275 °C (-30 to
525 °F)
Accuracy
Assumes ambient operating
temperature of 23 °C ±2 °C (73 °F
±4 °F)
Between 0 °C and 65 °C (32 - 150
°F): ±1 °C (± 2 °F) Below 0 °C (32
°F): ±1 °C (± 2 °F)
± 0.1 degree/degree Above 65 °C
(150 °F): ±1.5 % of reading
Response Time <500 ms after initial reading
Spectral response 8-14 microns
Emissivity
[1]
Pre-set for Food Service
Applications
Distance to Spot Size Optical
Resolution (D:S)
2.5:1 @ 90 % energy, typical
Typical Working Range (target
illumination)
25 mm to 250 mm (1" to 10")
Minimum Target Size 12 mm (0.5 ")
Illumination to IR Channel Offset 13 mm (0.52 ")
Probe
Temperature Range -40 °C to +200 °C (-40 °F to
+392 °F)
Accuracy
Assumes ambient operating
temperature of 23 °C ± 2 °C (73 °F
± 4 °F)
Between -5 °C and 65 °C (23 °F -
150 °F):
±0.5 °C (±1 °F)
Below -5 °C (23 °F): ±1 °C (± 2 °F)
Above 65 °C (150 °F): ±1 % of
reading