Fluke 50D Thermometer User Manual


 
50S & 50D Thermometer
Calibration Information
12
5. Allow the system to stabilize, then adjust T1 or T2 offset adjustments for a temperature display equal to
the reading of the lag bath reference thermometer.
6. Calibration of the thermometers is complete. Disconnect all test equipment from the instrument.
Lag bath
(room temperature
water)
Mercury
thermometer
(0.02
°C resolution)
Immersed to
approximately
same depth
Thermocouple wire
50S or 50D
Thermometer
_
+
zf06f.eps
Figure 7. Room Temperature Lag Bath