Fluke 1524 Thermometer User Manual


 
1523, 1524 Thermometer Readout
Specications
26
Ohms Measurement, ThermistorTable 9
Ohms Range
Accuracy ± Ω, 4 Wire
200 to 50 k ± (0.01 % + 0.5 )
50 k to 500 k
± (0.03 %)
Temperature Coefficient ( –10 °C to 13 °C , +33 °C to 60 °C):
0.002 %/°C + 0.1(0 to 50 k)
0.06 %/°C + 0.1(50 k to 500 k)
Excitation Current:
10 µA (0 to 50 k)
2 µA (50 k to 500 k)
Equivalent temperature accuracies derived from primary specications (, mV)
Temperature, ThermocouplesTable 10
Type Range
Measure
Accuracies (ITS-90)
B 600 °C to 800 °C
800 °C to 1000 °C
1000 °C to 1800 °C
0.85 °C
0.68 °C
0.57 °C
C 100 °C to 550 °C
550 °C to 2300 °C
0.32 °C
0.71 °C
E –200 °C to 0 °C
0 °C to 950 °C
0.52 °C
0.22 °C
J –200 °C to 0 °C
0 °C to 1200 °C
0.52 °C
0.23 °C
K –200 °C to 0 °C
0 °C to 1370 °C
0.61 °C
0.24 °C