Exergen 2000C Thermometer User Manual


 
Display Messages
A flickering Scn on display is visible during measurement.
At completion, releasing the button will display and lock
temperature on the screen for 30 seconds.
The target temperature measured is higher than 107.6°F
(42°C).
The target temperature measured is lower than 60°F
(15.5°C).
Temperature of the thermometer is higher than 104°F
(40°C). Let the instrument acclimatize for about 30 min-
utes in a cooler area in which it will be used.
The temperature of the thermometer is lower than 60°F
(15.5°C). Let instrument acclimatize for about 30 minutes
in a warmer area in which it will be used.
EMI/RFI (like static on a radio) protection is preventing a tem-
perature from being taken. Wait a minute and you should be
able to proceed. If not, reset by removing and replacing the
battery. Be sure battery is tightly connected. Call Customer
Service if error message reappears.
°C/°F Conversion
Remove the battery cover
and battery. Remove the clip
shown by the arrow.
Replace the clip in either the
°C position or the °F position
shown by the diagram.
Replace battery and battery cover.
14
tat2000instrev6.qxd 2/6/2006 9:35 AM Page 14