Excell Precision WEIGHING SCALES Scale User Manual


 
V1.1 ZSME300000011
20
EXCELL PRECISION CO., LTD.
(3) Data Format
Stable transmission ( ) / Continuous transmission ( )
Press the ¿ key to transmit ( )
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
kg/lb
HEAD 1 , HEAD 2 , DATA UNIT CR LF
HEAD 1 ( 2 BYTES ) HEAD 2 ( 2 BYTES )
OL - Overload , Under load
TR - TARE Mode
ST - Display is Stable NT - NET Mode
US - Display is Unstable GS - GROSS Mode
DATA (8 or 9 BYTES)
2D ( HEX ) = ( MINUS )
2E ( HEX ) = . ( DECIMAL POINT )
UNIT ( 2, 3 or 4 BYTES )
kg = 6B ( HEX ) ; 67 ( HEX )
lb = 6C ( HEX ) ; 62 ( HEX )
tl.T = 74 ( HEX ) ; 6C ( HEX ) ; 2E ( HEX ) ; 54 ( HEX )
hkg = 68 ( HEX ) ; 6B ( HEX ) ; 67 ( HEX )
Transmission example:
1. The gross weight (+0.876kg) shows as below, after stable: (in the tare mode)
S
T
, G
S
, +
0
. 8
7
6
k
G
0D
0A
HEAD 1 , HEAD 2 , DATA UNIT CR LF
2. The net weight (-1.568lb) shows as below without weight stability: (Not in the tare mode)
U
S
, N
T
, - 1
. 5
6
8
l b
0D
0A
HEAD 1 , HEAD 2, DATA UNIT CR LF
3. The net weight (+0.876kg) shows as below, after stable: (in the tare mode)
S
T
, T
R
, +
0
. 8
7
6
k
G
0D
0A
HEAD , HEAD , DATA UNIT C LF