Braun BP3550 Blood Pressure Monitor User Manual


 
12
What to do if ...
Problem Reason Solution
(flashing)
Batteries are flat.
Memory readings can be
called up.
Measurements are not
possible.
Insert new batteries.
e.g. Error codes 01-29:
Device may not be
correctly positioned on
the wrist.
Refit device and repeat
measurement as shown and
described under «Measuring».
e.g. Error codes 30 and
higher:
System error
Repeat measurement. If problem
persists, have device checked at a
Braun Service Centre.
Measurement
or inflation
does not start
instantly.
Right measuring position
was not reached.
Repeat measurement as shown
and described under «Measuring»
or deactivate the positioning
system (A).
Readings from
repeated mea-
surements differ
considerably.
Blood pressure is a
fluctuating value.
For healthy adults,
deviations of about 10 to
20 mmHg are possible.
none
Cuff is not properly fitted. Ensure that the appliance is fitted
as follows:
– on top of the slim side of the
forearm
– with the display on top
– with the battery compartment
facing the user’s body
– approx. 2 cm away from the
thumb root
Readings were not con-
sistently taken at heart
level.
For each measurement, proceed
exactly as shown and described
under «Measuring».
Talking, coughing,
laughing, moving etc.
when measuring will
influence the reading.
While measuring, relax, keep still,
do not move, or speak.
Prior to a
measurement,
the positioning
system (A) on the
display does not
work.
Positioning system (A) is
deactivated.
Refer to the chapter «Select mode»
to activate the positioning system
(A).
................
................
6084403_BP3550/60_S4-76 Seite 12 Montag, 5. Februar 2007 2:30 14