7
the measuring process when the probe is correctly
positioned, and lights up continuously when the
thermometer detects that an accurate measurement
has been taken.
4.
If the probe has not been positioned correctly in
the ear canal or was moved during the measuring
process, a sequence of short beeps will sound, the
«ExacTemp» light will go out and the display will
show an error message («POS» = position error).
5.
For the next measurement, eject the used probe
cover (push ejector (4)) and put on a new, clean
probe cover. The thermometer will turn on auto-
matically.
Wait for the ready signal. Fit the probe snuggly into
the ear canal, then push and release the Start button.
The Braun ThermoScan ear thermometer turns off
automatically after 60 seconds of inactivity. It can
also be turned off by pressing the «I/O» button for at
least three seconds. The display will shortly flash
«OFF» and steadily display the word «OFF» after
releasing the button.
Temperature taking hints
• A temperature measurement taken in the right ear
may differ from a measurement taken in the left
ear. Therefore, always take the temperature in the
same ear.
• The ear must be free from obstructions or excess
cerumen build-up in order to take an accurate
measurement.
• External factors may influence ear temperatures,
particularly when an individual has:
– been lying on one ear or the other
– had their ears covered
– been exposed to very hot or very cold temper-
atures, or
– been recently swimming or bathing.
In these cases, remove the individual from the
situation and wait 20 minutes prior to taking a
temperature.
• For persons wearing hearing aids or ear plugs,
remove the device and wait 20 minutes prior to
taking a temperature.
• Use the untreated ear if ear drops or other ear
medications have been placed in the ear canal.