4 - Running the Test
34
3. Select Perform post processing automatically at end of test execution to process test data
immediately after the completion of the test. Post processing is required to generate test reports
and to generate data for the Statistics and Probability displays. Depending on the length of the
test, the post processing of data may require considerable time. Alternatively, you can post
process the test data at a later time by opening the test file and pressing the Run Post processing
button.
4. Press the Run Test button to run the test.
Note The drop-down menu commands are not available while the test is running.
The Test window presents a summary of the test setup for the selected test type. The progress bar on the
screen indicates the remaining time until completion of the test. Pressing the Abort Test button lets you
abort the test at any time. All of the information up to the time that the test is aborted is saved in the data
file.
If Perform post processing automatically is checked, post processing will automatically be performed
on all test data even if the test is aborted.
The Mains control lets you control the output from the Agilent 6800-Series AC Power Source/Analyzer.
This software control acts the same as the Output On/Off button on the front panel of the unit. Use the
Mains menu to configure the Mains control. Pushbuttons appear in the mains control area when it is
active. When the control is inactive, the pushbuttons do not appear but are replaced by a round circle that
indicates the output state. A white circle indicates the output is off. A red circle indicates the output is
on. The Mains field on the menu bar at the top of the screen also indicates the present state of the output.
Note If you use the Output On/Off button on the front panel of the unit to turn the output on or
off, the Mains control does not reflect the changed condition on the output until you exit
and then re-enter the Test window. This action updates the Mains control display.
Configure the Mains control using the Mains menu. Note that the configuration of the Mains control
is saved when you exit the Agilent 14761A HFTS software. When you run the Agilent 14761A HFTS
software again, the unit will wake up with the output set according to the state in which you left the
Mains switch .
Auto On/Auto Off
The output of the Agilent 6800-Series AC Power
Source/Analyzer is automatically controlled by the Agilent
14761A HFTS software while pre-tests and tests are
running. No pushbuttons appear on the control. The pictured
control indicates that the output is presently off.
Auto On/Manual Off
The OFF pushbutton is active, letting the operator manually
turn the output of the Agilent 6800-Series AC Power
Source/Analyzer off from the Test window. The output is
never automatically removed from the equipment under test
if Auto On/Manual Off is selected. The pictured control
indicates that the output is presently off.