Agilent Technologies 6834B Pacemaker User Manual


 
Performance Tests
9
2
Performance Tests
Introduction
This chapter provides test procedures for checking the operation of an Agilent Model 6814B/6834B AC Power
Source/Analyzer or Agilent Model 6843A RTS. The required test equipment is specified and sample performance test
record sheets are included. Instructions are given for performing the tests using the front panel keypad.
Test Equipment Required
Table 2-1 lists the equipment required to perform the verification tests given in this chapter.
Table 2-1. Test Equipment Required
Equipment Characteristics Recommended Model
Digital Voltmeter* Resolution: 10nV @ 1V
Readout: 81/2 digits
Accuracy: <20ppm
Agilent 3458A
Current Monitor 0.01 ohms +/-200ppm 10W Guildline 7320/0.01
Audio Analyzer Input Voltage Range: 50mV to 300 V
Distortion Accuracy: +/-1 db 20Hz to 20KHz
Residual Distortion / Noise: -80db (0.0 1 %) 20Hz to 20KHz
Agilent8903A
Frequency Counter Accuracy @ 1 KHz < 0.001% Agilent5316B
Ratio Transformer* 30:1 ratio < 50ppm
Variable –Voltage
Transformer
Power: 3 phase 24 KVA
Range: 180-254V 47-63 Hz
360-440V 47-63 Hz
Superior Powerstat
1156DT-3Y
0-280 50A 24.2KVA
or equivalent
Load Resistors 6814B use 4 ea. 7.5 ohms > 1000W
6834B use 4 ea. 15 ohm > 1000W
6843A use 4 ea. 5 ohm > 1500W
GPIB Controller Full GPIB capabilities HP Series 200/300
* The 30:1 ratio transformer is not required for a valid MIL-STD 45662A test. The 30:1 ratio transformer is only required when a 4:1
test equipment to specification ratio is desired using a 3458A voltmeter in the 1000 volt range.
General Measurement Techniques
The following chapters provide the test procedures for verifying the Agilent 6814B, 6834B or 6843A's compliance with the
specifications listed in Appendix A of the Users Guide. Please refer to CALIBRATION PROCEDURE or
TROUBLESHOOTING if you observe out-of specification performance. The performance test specifications are listed in
the Performance Test Record Card at the end of this chapter. You can record the actual measured values in the columns