A&D UA-774 Blood Pressure Monitor User Manual


 
LifeSource Health Line (Toll-Free): 1-888-726-9966
Display
Symbol
Recommended
Action
Measurement is in progress,
remain as still as possible.
Take measurement again
and consult with your
physician.
Replace all batteries with
new ones when the
indicator blinks.
Try the measurement again.
Remain very still during the
measurement.
Fasten the cuff correctly,
and try the measurement
again.
Check for air leaks along
the tube and around the
air socket.
Refasten the cuff and
retake measurement
Try the measurement again.
Remain very still during
the measurement.
Measurement is in progress,
remain as still as possible.
Condition/Cause
The indicator while
measurement is in progress.
It blinks while detecting the
pulse.
An irregular heartbeat or
body movement may have
occurred. Refer to page E-10
for more information on
irregular heartbeats.
Previous measurement
stored in MEMORY.
The battery power indicator
during measurement.
The batteries are low when
it blinks.
Unstable blood pressure due
to excessive body
movement.
The systolic and diastolic
values are within 10 mmHg
of each other.
The pressure value did not
increase during inflation.
The cuff is not fastened
correctly.
The pulse is not detected
correctly.
Pressure Indicator
WHAT DISPLAY SYMBOLS MEAN
?
E-2
?
Battery Full
Battery Low
PUL. DISPLAY ERROR