A&D TM-2655P Blood Pressure Monitor User Manual


 
20
9-2 Replacing the Fuses
Replace the fuses as follows:
1. Remove the caps of both fuse holders located on the rear
panel.
2. Replace both fuses with new ones.
3. Re-install the fuse holder caps.
Fig.19 Fuse holders
Note
Use only fuses of the proper type, voltage and current rating as specified on the rear panel, and that
conform to IEC60127.
9-3 Checking the Counter
The TM-2655/TM-2655P is equipped with a counter function, which indicates how many times the device
has measured blood pressure. The counter data remains in memory even after the power is turned off.
Use the counter function to check the frequency of the device use or as the guideline for a periodic
cleaning
Displaying the counter
Press the COUNTER switch. The counter data appears in the
SYSTOLIC and DIASTOLIC displays for about 4 seconds.
Note
Do not keep the COUNTER switch pressed. It will reset the counter
data to zero.
Resetting the counter
Hold down the COUNTER switch for 4 seconds or more. The
counter data will be reset to zero.
9-4 Error Codes
Table 9 Error codes
Error code Description
pe
The printer paper has run out. Install a new roll of printer paper.
HU
The printer head is up. Lower the lever and lock the printer paper in position.
err
A measurement error has occurred. Refer to “10. Troubleshooting”.
Fig.20 COUNTER switch