A&D TM-2655P Blood Pressure Monitor User Manual


 
5
1-3 Specifications
Table 1 Specifications
Performance specifications
General
Power supply 120VAC 60 Hz (For USA area)
230VAC 50 Hz (For Europe area)
(factory-preset)
Power consumption Approx. 40VA
Protection against electrical shock
Class I , Type B
Display LED
EMC EN 60601-1-2
Blood pressure measurement
Blood pressure measuring method Oscillometric method
Pressure detection method Capacitance type pressure transducer
Pressure range 0 – 300 mmHg
Accuracy
Pressure: ±3 mmHg
Pulse rate: ±5%
Measurement range Blood pressure: 10 – 280 mmHg
Pulse rate: 30 – 200 bpm
Pressurizing method Micro pump
Air pressure control method Rubber valve, ceramic valve
Rapid air exhaust system Electromagnetic valve
Safety mechanism The electromagnetic valve is released when
approx. 320 mmHg or greater is detected.
Communications functions Serial output: RS232C level
Environment specifications
Operating temperature and humidity
10 to 40°C, 85%RH or less, non condensing
Storage temperature and humidity
-20 to 60°C, 95%RH or less, non condensing
Physical specifications
Overall dimensions 245(W) X 320(H) X 390(D) mm
Weight Approx. 9 kg