A&D HL-3000WP Scale User Manual


 
4
WEIGHING
Before weighing
Automatic Power-Off Function
If the scale is left ON and the stable indicator is displayed, the automatic
power-off function turns the power off after approximately 5 minutes
(factory setting). To toggle the function disabled/enabled, press the
[ON/OFF] key while pressing the [RE-ZERO] key. The scale stores the
setting and it will be shown by pressing both the [RE-ZERO] and
[UNITS] keys in the weighing mode. (See “CALIBRATION” about the
[UNITS] key.)
poff0e Function disabled. poff1e Function enabled.
Error message
eeeee Overload
----- The scale zero is out of range
ealba Low battery
Calibration may be required before weighing
Read “CALIBRATION” first and if necessary, calibrate your scale for
accurate weighing.
Weighing procedures
1. Turn on the scale.
Press the [ON/OFF] key to turn on the scale. Then the scale displays all
segments for a few seconds and then 0 will be displayed.
2. Start weighing
If you do not use a container for weighing,
Verify the reading is 0 . If not, press the [RE-ZERO] key to display
0 . Place objects on the weighing platform to weigh. When the reading
becomes stable, the stable indicator is displayed.
If you use a container for weighing,
Place an empty container on the platform.
Wait for the stability indicator to be displayed and press the [RE-ZERO]
key. Place the objects to be weighed in the container.
Select the weighing unit, --- Only for U.S.A. models ---
If you use the U.S.A. model, press the [UNITS] key to select weighing unit.
Normally scales power up in “lb” or “oz” when shipped from the factory. To
power up in other units, turn the scale off. Press and hold the [UNITS] key,
then press the [ON/OFF].key The scale will start with the next unit.
Note) HL-300WP: ozÆgÆozÆ
HL-1000WP/3000WP/3000LWP: lbÆozÆlb-ozÆgÆlbÆ