A&D HC-3Ki Scale User Manual


 
44
The HC-i is designated as DCE (Data Communication Equipment).
13-3. Data Output Mode
The Data Output Modes and Parameters are set by F-Functions in F-06-X as
described in the “9-2. F-Functions”
To control the scale using commands form an external device, see “13-6. Command
Mode”.
Refer to “13-6. Command Mode” about the output data format.
Data Output Mode (f-06-01)
Key Mode (f-06-01="0")
When the weight display is stable, data is sent by pressing the PRINT key. The
count display will blink when the data has been sent.
Stream Mode (f-06-01="1")
Data is sent continuously. The data-update rate is approximately 10 times per second for
f-06-03="2". For f-06-3="0" or "1", the interval between continuous data is
approximately 2 seconds.
Auto-print Mode A (f-06-01="2")
Data is sent if the weight display is stable at +5d (weighing display division) and
above. The next transmission can not occur until after the weight display falls below
+5d.
Auto-print Mode B (f-06-01="3")
Data is sent if the weight display is stable at ±5d (weighing display division) and
above/below. The next transmission can not occur until after the weight display falls
between –5d and +5d.
To use with the UFC format, refer to "13-7. Using UFC (Universal Flex Coms)
Function"
Data to be Sent (f-06-02)
Select which data to be sent by keying in a 0 or 1 for the data: ID No., PCS
(count), weight or unit weight.
Example: Key in 1 1 0 0 to
display 1100, this setting would
send only the ID number and the
count.
Data Format (f-06-03)
Format for AD-8121 MODE 1 or 2. (f-06-03="0")
Format for AD-8121 MODE 3. (f-06-03="1")
Format for general apparatuses, computers, etc. (f-06-03="2")
0’ not to send data
1’ to send data
8f-06-02
8
8Clear
8
8
ai11000
00
ID No.
PCS (count)
subtr
acts